A recall has been issued for certain Phillips Respironics ventilators, BiPAP, and CPAP machines. The recall was issued because of black debris in the air pathway of the units. The FDA is analyzing reports related to these devices that range from 2009 to 2021. Below is the information from the FDA and a list of the affected products. To register a recalled product visit Register your device(s) on Philips Respironics’ recall website

The U.S. Food and Drug Administration (FDA) is alerting people who use Philips Respironics ventilators, BiPAP, and CPAP machines and their health care providers that Philips Respironics has recalled certain devices due to potential health risks. The polyester-based polyurethane (PE-PUR) sound abatement foam, which is used to reduce sound and vibration in these affected devices, may break down and potentially enter the device’s air pathway. If this occurs, black debris from the foam or certain chemicals released into the device’s air pathway may be inhaled or swallowed by the person using the device.

Affected CPAP & BiPAP Devices

  • E30
  • Dreamstation ASV
  • Dreamstation ST, AVAPS
  • SystemOne ASV4
  • C-Series ASV
  • C-Series S/T and AVAPS
  • OmniLab Advanced+
  • SystemOne (Q-Series)
  • DreamStation
  • DreamStation Go
  • Dorma 400
  • Dorma 500
  • REMstar SE Auto

Affected Ventilators

  • Trilogy 100
  • Trilogy 200
  • Garbin Plus, Aeris, LifeVent
  • A-Series BiPAP Hybrid A30 (not marketed in US)
  • A-Series BiPAP V30 Auto
  • A-Series BiPAP A40
  • A-Series BiPAP A30

Risk Factors

Polyester-based polyurethane (PE-PUR) is a sound abatement foam used to reduce sound and vibration in these devices and other medical equipment. The PE-PUR foam in the affected Philips Respironics CPAP, BiPAP, and ventilator devices may:

  • Break down (degrade) into particles which may enter the device’s air pathway and be inhaled or swallowed by the user
  • Release certain chemicals into the device’s air pathway, which may be inhaled

These issues can result in serious injury, which can be life-threatening, cause permanent impairment, and require medical intervention to prevent permanent damage.

To date, Philips Respironics has received several complaints about the presence of black debris/particles within the device’s air pathway. Philips Respironics also has received reports of headache, upper airway irritation, cough, chest pressure, and sinus infection, which may be related to this issue, though the cause of the symptoms cannot be definitively linked.

The potential risks of particulate exposure include irritation to the skin, eye, and respiratory tract, inflammatory response, headache, asthma, and toxic or carcinogenic effects to organs, such as kidneys and liver.

The potential risks of exposure to chemicals released into the device’s air pathway from the PE-PUR foam include headache; dizziness; irritation in the eyes, nose, respiratory tract, and skin; hypersensitivity; nausea/vomiting; and toxic and carcinogenic effects.

The foam degradation may be exacerbated by high heat and high humidity environments, and by use of unapproved cleaning methods, such as ozone.

To date, there have been no reports of death as a result of these issues.